The IQT®-XLM unit is specified equipment for testing fuels according to:
Standard IQT®-XLM System Inclusions:
Suitable for 50 Hz and 60 Hz electrical supplies
Automated control of injection nozzle and combustion sensor temperature (formerly requiring the TALM option) is standard equipment
Electronic or mechanical pressure regulation system for compressed air and nitrogen
|
Motor Method (ASTM D6890) |
---|---|
Approved Reporting Range |
31.5 to 75.1 DCN (working range of ASTM D6890 per ASTM D6300) |
International Fuel Specifications / Types Approved for Cetane Measurement |
ASTM D975 (petroleum diesel), ASTM D6751 (biodiesel) |
Warmup Time |
Approximately 30 minutes to reach operating temperature from room temperature |
Test Time Per Sample |
Approximately 17 minutes (includes 15 warmup injections and 32 injections for analysis) |
Reported Test Result |
Average ignition delay and DCN values of final 32 injections |
Test Result Data Base |
All test reports and sensor data recorded and stored during the lifetime of the instrument |
Sample Volume |
35 ml for cleaning, 30 ml for test run |
Compressed Gas Requirements |
Compressed air (19.9-21.9% O2) at 21.7 bar (310 psi) |
Reference Materials |
n-heptane 99.5%, octane engine grade (calibration reference material, check standard) |
Test Sample Vessels |
Removable stainless steel fuel reservoirs (26 ml, 55 ml, or 95 ml) |
Electrical Connection |
20A, 208-240 VAC, 50/60 Hz |
Operating Conditions |
Ambient temperature: 18 to 32° C, exhaust system ventilation: ~100 cfm |
Approximately 66.8 cm (26.3 in) H x 118.1 cm (46.5 in) W x 56.4 cm (22.2 in) D, approximately 170 kg (375 lbs)