Gravimetric Blending System (GBS)
Offering unmatched precision of +/- 0.2% for the gravimetric blending of primary, secondary, and standard reference fuels.
Elevating your online blending operations, the Continuous Testing System (CTS) seamlessly integrates with cutting-edge XCP TECHNOLOGY, enabling your Octane testing operation to function autonomously 24/7 by connecting one or more engines to your octane blending system.
As with all CFR products, the CTS is meticulously designed, manufactured, and tested to deliver a cohesive and dependable solution for your operations. CFR provides the engine, parts, accessories, instrumentation, and control system, ensuring they work together with unmatched efficiency.
Download Product Details
Introducing the CFR Continuous Testing System (CTS) for testing fuels according to:
Touch screen HMI with industrial PC
Microsoft Windows operating system
Modbus register for SCADA connectivity
Adjustable fuel bowl height
Lamba sensor for Air/Fuel ratio
Pumps and solenoids for delivering and draining the fuel
LIMS connectivity
Bolt-on configuration on existing units or sold as a new unit
Power Supply: 120V, 1Ph, 50/60 Hz
|
Motor Method (ASTM D2700) |
Research Method (ASTM D2699) |
---|---|---|
RPM |
900 +/- 1% |
600 +/- 1% |
Timing |
Variable based on cylinder height |
13° BTDC |
Water Jacket Temperature |
100 °C +/- 1.5 °C (212 °F +/- 3 °F) |
100 °C +/- 1.5 °C (212 °F +/- 3 °F) |
Oil Temperature |
57 °C +/- 8 °C (135 +/- 15 °F) |
57 °C +/- 8 °C (135 +/- 15 °F) |
Oil Pressure |
172 kPa-207 kPa (25 psi-30 psi) |
172 kPa-207 kPa (25 psi-30 psi) |
Vacuum |
25-150 mm H2O (1-6 in H2O) |
25-150 mm H2O (1-6 in H2O) |
Intake Temperature |
38 °C +/- 2.8 °C (100 °F +/- 5 °F) |
Standard Temperature based on Barometric Pressure |
Mixture Temperature |
Standard 149 °C (300 °F+/-2F) |
N/A |
Call +1 262.501.5998 or complete this form and we'll get back to you quickly!